Agilent Technologies DP235 Guía de usuario Pagina 427

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 628
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 426
427
Troubleshooting the Logic Analyzer
Analysis Probe Problems
Slow clock
If you have the analysis probe hooked up and running and observe a
slow clock or no activity from the interface board, the +5 V supply
coming from the analyzer may not be getting to the interface board.
To check the +5 V supply coming from the analyzer, disconnect one of the
logic analyzer cables from the analysis probe and measure across pins 1
and 2 or pins 39 and 40.
If +5 V is not present, check the internal analysis probe fuse or current
limiting circuit on the logic analyzer. For information on checking this fuse
or circuit, refer to the 1670G-Series Logic Analyzers Service Guide.
If +5 V is present and the cable connection to the analysis probe appears
sound, contact your nearest Agilent Technologies Sales Office for
information on servicing the board.
Vista de pagina 426
1 2 ... 422 423 424 425 426 427 428 429 430 431 432 ... 627 628

Comentarios a estos manuales

Sin comentarios