Specifications
General Information
Measurement and display functions
Displayed waveforms. 24 lines maximum, with scrolling across 96
waveforms.
Measurement functions
Run/Stop functions. Run starts acquisition of data in specified trace
mode.
Stop. In single trace mode or the first run of a repetitive acquisition,
Stop halts acquisition and displays the current acquisition data. For
subsequent runs in repetitive mode, Stop halts acquisition of data and
does not change the current display.
Trace mode. Single mode acquires data once per trace specification.
Repetitive mode repeats single mode acquisitions until Stop is pressed
or until the time interval between two specified patterns is less than or
greater than a specified value, or within or not within a specified range.
Indicators
Activity indicators. Provided in the Configuration and Format
menus for identifying high, low, or changing states on the inputs.
Markers. Two markers (X and O) are shown as vertical dashed lines
on the display.
Trigger. Displayed as a vertical dashed line in the Timing Waveform
display and as line0 in the State Listing display.