Logic Analyzer Reference
The Analyzer Format Menu
The Analyzer Format Menu
Pod threshold field
The pod threshold field is used to set a voltage level that the data must
reach before the analyzer recognizes and displays it as a change in logic
levels. Threshold levels apply to single pods, and cover both data and
clock channels.
TTL. When TTL is selected, the threshold level is +1.5 volts.
ECL. When ECL is selected, the threshold level is -1.3 volts.
User. When User is selected as the threshold level, the data signals
must reach a user selectable value. The range of this value is between -
6.0 volts to +6.0 volts in 50-mV increments.