Agilent Technologies DP235 Guía de usuario Pagina 202

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 628
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 201
202
Using the Pattern Generator
Building Test Vectors and Functions
To build an initialization sequence
Use the INIT SEQUENCE to place the system under test into a known
initialization state. Default start and end program vectors are marked
INIT SEQUENCE START and INIT SEQUENCE END. During a
repetitive run, the initialization sequence is only executed the first time
the program is run. The main sequence then loops repetitively.
1 From the Sequence menu, use the knob to highlight INIT
SEQUENCE START.
2 Select the Insert field once for each new line of vectors you want
inserted into the initialization sequence.
These new lines of vectors provide fields to place data into for each
label.
3 Select the data field, then type in a data value.
If you are adding data in many fields, use the Autoroll feature. Refer to
"To use Autoroll" on page 217.