Logic Analyzer Reference
Probing
Data Cable Characteristics Without a Data Pod
The 1670G-series, with the pattern generator option, data cables
without a data pod provide an ECL-terminated (1 KW to -5.2 V)
differential signal. These are usable when received by a differential
receiver, preferably with a 100-ohm termination across the lines. These
signals should not be used single ended due to the slow fall time and
shifted voltage threshold (they are not ECL compatible).