Triggering Commands 9
U2761A Programmer’s Reference 99
TRIGger[:SEQuence]:SLOPe
Syntax
TRIGger[:SEQuence]:SLOPe {POSitive|NEGative}
This command sets the rising or falling edge for the trigger signal on the
Trig In connector to start the sweep.
TRIGger[:SEQuence]:SLOPe?
This query returns the selected trigger slope as POS or NEG.
Parameter
Examples
The following command sets the trigger slope to Positive.
TRIG:SLOP POS
The following query returns the trigger slope.
TRIG:SLOP?
Typical response:
POS
Item Type Range of values Default value
slope Discrete POSitive|NEGative POSitive
TRIG :SEQ space:SLOP POS
NEG
?