Agilent Technologies AGILENT U2761A Manual de servicio Pagina 60

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 169
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 59
4 Output Configuration Commands
52 U2761A Programmer’s Reference
[SOURce:]FUNCtion:RAMP:SYMMetry
Syntax
[SOURce:]FUNCtion:RAMP:SYMMetry <percent>
This command sets the symmetry percentage for the Ramp wave.
[SOURce:]FUNCtion:RAMP:SYMMetry?
This query returns the current symmetry setting in %.
Parameter
Examples
The following command sets the symmetry to 50%.
FUNC:RAMP:SYMM 50
The following query returns the symmetry.
FUNC:RAMP:SYMM?
Typical response:
+5.0000000000000E+01
Item Type Range of values Default value
percent Numeric 0% to 100% 100%
SOUR : FUNC space
?
:RAMP :SYMM percent
NOTE
Setting the symmetry to 50% will produce a Triangle wave, and 0% for a Negative Ramp
wave.
Vista de pagina 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 168 169

Comentarios a estos manuales

Sin comentarios