CALibration Commands 14
U2761A Programmer’s Reference 133
CALibration:VALue
Syntax
CALibration:VALue <value>
This command specifies the measured value of the known calibration
signal depending on the calibration steps.
CALibration:VALue?
This query returns the calibration value.
Examples
The following command sets the value to 1×10
7
.
CAL:VAL 1E+7
The following query returns the calibration value.
CAL:VAL?
Typical response:
+1.0000000000000E+07
CAL space
?
value:VAL
NOTE
Refer to the Agilent U2761A USB Modular Function/Arbitrary Waveform Generator Service
Guide for the range of values of each calibration step.