9 Triggering Commands
102 U2761A Programmer’s Reference
OUTPut:TRIGger[:STATe]
Syntax
OUTPut:TRIGger[:STATe] {0|OFF|1|ON}
This command disables or enables thetrigger out signal on the Trig Out
connector. When enabled, a TTL- compatible Square waveform with the
specified edge is output from the Trig Out connector at the beginning of
the sweep.
OUTPut:TRIGger[:STATe]?
This query returns the trigger output state as 0 if the “trigger out” signal
is disabled, or 1 if the signal is enabled.
Parameter
Examples
The following command enables the trigger output.
OUTP:TRIG ON
The following query returns the trigger output state.
OUTP:TRIG?
Typical response:
1
Item Type Range of values Default value
state Boolean 0|OFF|1|ON 0
OUTP :TRIG space
?
0|OFF
1|ON
:STAT