CALibration Commands 14
U2761A Programmer’s Reference 131
CALibration:SECure:CODE
Syntax
CALibration:SECure:CODE <new code>
This command inputs a new security code. To change the security code,
you must first unsecure the U2761A using the old security code, and then
enter a new code. The security code is stored in nonvolatile memory.
Parameter
Remark
The first character of the code must always be a letter (A to Z).
Example
The following command renews the security code of the U2761A.
CAL:SEC:CODE U2761A
Item Type Range of values Default value
code String Up to 12 alphanumeric characters U2761A
CAL space:SEC :CODE new code