7 Press the Define Automask softkey.
8 Turn the knob closest to the Cursors key until the number highlighted
under the
Mask softkey is 1.
9 Press the Create Mask softkey.
A mask is generated from the signal on the screen. The mask consists of two
lines: the maximum limit line and the minimum limit line.
10 Press the Previous Menu softkey, then press the Define Sequence
softkey.
11 Press the Step softkey or turn the knob closest to the Cursors key
until the number highlighted under
Step is 1.
12 Press the Save Setup softkey.
13 Press the Use Mask softkey. Then press the Use Mask softkey, or
turn the knob closest to the Cursors key, until the number highlighted
is 1.
Quick Start
To set up the Test Automation Module
13