To set a step or mask to default values
1 Press Print/Utility twice.
2 Press the Sequence Admin softkey.
3 Press the Copy Step/Mask softkey.
4 Toggle the Type softkey to select a step or a mask.
5 Press the Dest softkey.
6 Press the Dest softkey, or turn the knob closest to the Cursors key,
to select the number of the step or mask.
7 Press the Default Dest softkey to set the step or mask to its default
values.
The default settings of a step are:
Mask None Branching
Test Mode Pass Next
Adjust Mode On Fail Min Next
Permit All Fail Max Next
Fail when Out Outputs
Message 1 PASS
Label ** Test #XX
**
2 FAIL
Pass PASS 3 OFF
Fail Min FAIL 4 OFF
Fail Max FAIL 5 OFF
Figure 11 shows the default settings for a mask.
Operating the Test Automation Module
To set a step or mask to default values
54