2 Press Print/Utility .
If the Previous Menu softkey appears on the rightmost softkey,
press
Print/Utility again.
When the module is installed, an additional softkey appears at the second
softkey from the left. For the HP 54655A, the label is HP-IB Menu. For the
HP 54656A, the label is RS-232 Menu.
3 Press the HP-IB Menu softkey or the RS-232 Menu softkey.
For HP 54655A, a softkey menu with five choices appears. Four of them are
related to the configuration of the HP-IB interface.
Connect to Selects the device connected to the oscilloscope.
Selections include HP plotter, HP printer, and computer.
Factors Selects whether the printer or plotter output contains the
information concerning the basic settings of the oscilloscope.
Resolution Selects the resolution of the printer or plotter output.
Address Selects the HP-IB address of the instrument. Valid addresses
range from 0 to 30. To change the address, press the softkey or turn the
knob closest to the Cursors key.
Previous Menu Returns the oscilloscope to the previous softkey menu.
For HP 54656A, a softkey menu with six choices appears. Five of them are
related to the configuration of the RS-232 interface.
Connect to Selects the device that the oscilloscope is connected to.
Selections include HP plotter, HP printer, Epson compatible printer, and
computer.
Factors Selects whether the printer or plotter output contains the
information concerning the basic settings of the oscilloscope.
Resolution Selects the resolution of the printer or plotter output.
Baud Rate Selects the baud rate of the RS-232 port. Valid baud rates
are 1200, 2400, 9600, and 19200.
Handshake Selects the handshaking method used in the RS-232. DTR
refers to the hardware handshaking using the DTR signal line. XON refers
to the software handshaking using the XON/XOFF protocol.
Previous Menu Returns the oscilloscope to the previous softkey menu.
Operating the Test Automation Module
To configure the interface
18