Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 173
:STATus
(see page 750)
Query Syntax
:STATus? <source>
<source> ::= {CHANnel<n> | FUNCtion | MATH | SBUS}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :STATus? query reports whether the channel, function, or serial
decode bus specified by <source> is displayed.
Return Format
<value><NL>
<value> ::= {1 | 0}
See Also "Introduction to Root (:) Commands" on page 138
":BLANk" on page 144
":CHANnel<n>:DISPlay" on page 208
":FUNCtion:DISPlay" on page 247
":SBUS:DISPlay" on page 401
":VIEW" on page 176
NOTE
MATH is an alias for FUNCtion.