Agilent Technologies InfiniiVision 5000 Series Manual de servicio Pagina 543

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 884
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 542
Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 543
:TRIGger:M1553:RTA
(see page 750)
Command Syntax
:TRIGger:M1553:RTA <value>
<value> ::= 5-bit integer in decimal, <nondecimal>, or
<string> from 0-31
<nondecimal> ::= #Hnn where n ::= {0,..,9|A,..,F}
<string> ::= "0xnn" where n::= {0,..,9|A,..,F}
The :TRIGger:M1553:RTA command sets the Remote Terminal Address
(RTA) to trigger on if the trigger type has been set to RTA using the
:TRIG:M1553:TYPE command.
Query Syntax
:TRIGger:M1553:RTA?
The :TRIGger:M1553:RTA? query returns the current TV trigger line
number setting.
Return Format
<value><NL> in nondecimal format
See Also "Introduction to :TRIGger Commands" on page 440
":TRIGger:M1553:TYPE" on page 546
Vista de pagina 542
1 2 ... 538 539 540 541 542 543 544 545 546 547 548 ... 883 884

Comentarios a estos manuales

Sin comentarios