Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 229
:DISPlay:LABList
(see page 750)
Command Syntax
:DISPlay:LABList <binary block data>
<binary block> ::= an ordered list of up to 75 labels, a maximum of 10
characters each, separated by newline characters.
The :DISPlay:LABList command adds labels to the label list. Labels are
added in alphabetical order.
Query Syntax
:DISPlay:LABList?
The :DISPlay:LABList? query returns the alphabetical label list.
Return Format
<binary block><NL>
<binary block> ::= an ordered list of up to 75 labels, a maximum of 10
characters each, separated by newline characters.
See Also "Introduction to :DISPlay Commands" on page 223
":DISPlay:LABel" on page 228
":CHANnel<n>:LABel" on page 211
NOTE
Labels that begin with the same alphabetic base string followed by decimal digits are
considered duplicate labels. Duplicate labels are not added to the label list. For example, if
label "A0" is in the list and you try to add a new label called "A123456789", the new label is
not added.