7-8
Bit Error Test Commands (Measure Subsystem)
FERasure:IRATio
FERasure:IRATio
Description Sets Frame ERasure Intermediate RATio MEASurement attributes. Queries the Frame
ERasure Intermediate RATio
GPIB units are % (PCT), PPM; default unit is PPM.
Display units are % (PCT), PPM; default unit is PPM.
NOTE This can only be queried when in the state: TRIGger:BETest:MODE ‘RUN’
Syntax MEASure:BETest:FERasure:IRATio<n>?
MEASure:BETest:FERasure:IRATio<n>[:MM] | [:AVG]
Options where <n> = 1..4.
Refer to Appendices D and F.
FERasure:RATio
Description Sets Frame ERasure RATio MEASurement attributes. Queries the Frame ERasure
RATio(completed).
GPIB units are % (PCT), PPM; default unit is PPM.
Display units are % (PCT), PPM; default unit is PPM.
NOTE This can only be queried when in the state: TRIGger:BETest:MODE ‘RUN’
Syntax MEASure:BETest:FERasure:RATio<n>?
MEASure:BETest:FERasure:RATio<n>[:MM] | [:AVG]
Options where <n> = 1..4.
Refer to Appendices D and F.
FLSelect
Description Selects/queries the Frame Loss Selected Bit Error Test measurement to display (Count or
Ratio) for the given Frame Loss TYpe (FE or CRC).
Syntax MEASure:BETest:FLSelect<n>?
MEASure:BETest:FLSelect<n> <string>
Options ‘COUNT’ | ‘RATIO’
where <n> = 1..4.