10-10
Configure Subsystem
OPERation:AUTO
OPERation:AUTO
Description Enables several auto-ranging routines, providing automatic adjustment of the affected
settings. Turns the RF Analyzer attenuator hold setting to AUTO.
(SANalyzer:ATTenuator:MODE ’AUTO’)
Turns the AF Analyzer gain cntl to AUTO.
(AFANalyzer:RANGing ’AUTO’)
Syntax CONFigure:OPERation:AUTO
Options Not applicable.
OPERation:HOLD
Description Disables several auto-ranging routines, requiring manual adjustment of the affected
settings.
Turns the RF Analyzer attenuator hold setting to HOLD.
(SANalyzer:ATTenuator:MODE ’HOLD’)
Turns the AF Analyzer gain cntl to HOLD.
(AFANalyzer:RANGing ’HOLD’)
Syntax CONFigure:OPERation:HOLD
Options Not applicable.
PRINt:ADDRess
Description Sets/queries the GPIB ADDRess of the PRINter connected.
Syntax CONFigure:PRINt:ADDRess?
CONFigure:PRINt:ADDRess <integer> | [:INUM]
Options Refer to Appendix A.