Vista de pagina 261
26-6
OSCilloscope Subsystem
TRIGger:MODE
TRIGger:MODE
Description Selects/queries how measurements are armed to accept a trigger.
IMPORTANT This command will set the trigger mode when in Local mode, it is overridden by
TRIGger:MODE:RETRigger REPetitive | SINGle when in Remote mode.
Syntax OSCilloscope:TRIGger:MODE?
OSCilloscope:TRIGger:MODE <string>
Options ’CONT’ | ’SINGLE’
Where;
CONT means that the oscilloscope is continuously armed to accept a trigger.
SINGLE means that the oscilloscope is armed to accept a trigger each time that
TRIGger:RESet is selected.
TRIGger:PRETrigger
Description Sets/queries the PRETrigger value. This is the number of divisions previous to the trigger
point.
Syntax OSCilloscope:TRIGger:PRETrigger?
OSCilloscope:TRIGger:PRETrigger <real> | [:INUM]
Options Refer to Appendix A.
TRIGger:RESet
Description Arms a measurement when TRIGger:MODE ‘SINGLE’ is selected or when
TRIGger:MODE:RETRigger SINGle is selected.
Syntax OSCilloscope:TRIGger:RESet
Options Not Applicable.
Vista de pagina 261