30-3
PULSe On/Off Ratio Subsystem
MARKer[:POSition]:FALL
MARKer[:POSition]:FALL
Description Sets/queries the PULSe MARKer FALL trace POSition setting. The value is given in units
of divisions from the left side of the trace (0 to 10 divisions).
Syntax PULSe:MARKer[:POSition]:FALL?
PULSe:MARKer[:POSition]:FALL <real> | [:FNUM]
Options Refer to Appendix B.
MARKer[:POSition]:RISE
Description Sets/queries the PULSe MARKer RISE trace POSition setting. The value is given in units
of divisions from the left side of the trace (0 to 10 divisions).
Syntax PULSe:MARKer[:POSition]:RISE?
PULSe:MARKer[:POSition]:RISE <real> | [:FNUM]
Options Refer to Appendix B.
MARKer:OPOSition:FALL
Description Sets/queries the PULSe Off POSition FALL setting. This is the time (relative to the center
of the last bit) that the amplitude on the amplitude envelope will be measured. The range
is 0.0 us to +56.0 us.
GPIB units are seconds (S), bit periods (T).
default unit is seconds (S),
default display unit is US (micro-second).
Syntax PULSe:MARKer:OPOSition:FALL?
PULSe:MARKer:OPOSition:FALL <real> | [:FNUM]
Options Refer to Appendix B.