Agilent Technologies N9360A Manual de usuario Pagina 134

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 354
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 133
5-16 N9360A Multi UE Tester GSM User Manual
5 Screen Reference
Automatic Test
Feature of Automatic Test
The test flow set on the [Configuration: Test Condition] screen
is automatically executed.
Setting for Automatic Test in Configuration Mode
Loss:
The path loss values for the radio systems can be set on the
[Configuration] screen to correct the RF attenuation
between the mobile phone and the Tester.
ATT In/Out:
The path loss values for each traffic channel can be set on the
[Configuration: Test Sequence] screen to correct the RF tests
results at each traffic channel in each radio system.
Sequence:
Up to two test sequences can be set on the [Configuration:
Test Sequence] screen.
Test S I M
Insert the Test SIM provided by Agilent in the mobile phone
before performing any test because the Tester will not be able to
perform measurements with a SIM that is provided by cell
phone operator or other Test SIMs.
Stand-by Screen
When the Automatic Test is pressed on the [Initial] screen, the
[Automatic Test: Stand-by] simplified screen is displayed. Then,
any of the [Automatic Test: Stand-by] simplified screen
(Figure 5-10), the [Automatic Test: Stand-by] detailed screen
NOTE
The default state at power on is the last setting stored in the internal
memory of the Tester before power off.
For numeric value entry, the changing magnification softkey is available.
Refer to Storing Numeric Values on page 8 and Changing Magnification
Softkey on page 13.
Vista de pagina 133
1 2 ... 129 130 131 132 133 134 135 136 137 138 139 ... 353 354

Comentarios a estos manuales

Sin comentarios