4-28 N9360A Multi UE Tester GSM User Manual
4 Operating Procedures
Step 14
Press the Start softkey to start a test. A screen status changes
from Stand-by to Measuring.
Step 15
Turn the mobile phone on. Wait for the mobile phone to camp
on, and P is shown at the Location Update step.
Step 16
At the MS Call step, make a call from the mobile phone. Dial an
arbitrary number and press the Off Hook button on the mobile
phone.
Step 17
At the Talk step, check the quality of loop back voice and press
the Pass or the Fail softkey according to its result.
Figure 4-19 [Automatic Test: Stand-by] Value Screen