Screen Reference 5
N9360A Multi UE Tester GSM User Manual 5-129
Trigger Sing/Cont: Selects the trigger mode from Sing (single)
and
Cont (continuous).
The
Sing mode setting provides single
measurement.
The
Cont mode setting provides
continuous measurement.
To start continuous measurement on the
[Measuring] screen, change the trigger
mode from
Sing to Cont.
To terminate continuous measurement on
the [Measuring] screen, change the trigger
mode from
Cont to Sing.
This softkey is effective only for changing
the trigger mode on the [Stand-by] screen.
Trigger: Starts single measurement. This softkey is inactivated
on the [Stand-by] screen.
Return: Returns to the previous screen.
Screen Field
Set the input fields of the following parameters to execute a new
test.
Table 5-54 [Manual Test] EGPRS Mode Sensitivity Screen Input Field
Input Field Description
Procedure Displays a list of parameter file names stored in the Tester.
Select a file to recall for a test.
If the Tester is using a parameter file, the file name is displayed in this field. If the Tester contains
parameter files, a list of those filename is displayed in this field for selection.
This field cannot be changed on a [Measuring] screen.
Radio System Selects a radio system from GSM850, GSM900, DCS1800 and PCS1900.
This field cannot be changed on the [Measuring] screen.
BCCH
*
Sets the broadcast control channel number.
The allowable range is as follows:
GSM850: 128 to 251
GSM900: 0 to 124 and 955 to 1023
DCS1800: 512 to 885
PCS1900: 512 to 810
This field cannot be changed on the [Measuring] screen.