4-20 N9360A Multi UE Tester GSM User Manual
4 Operating Procedures
Step 3
Set the Loss field to
On with the CURSOR CONTROL knob.
Step 4
Enter the appropriate loss values, depending on the radio
systems, in the
RF In and RF Out fields with the CURSOR
CONTROL knob. These loss values are used all through the test
flow.
Step 5
Press the Return softkey to return to the [Initial] screen.
Entering Channel Attenuations on the Configuration: Test Sequence Screen
Figure 4-12 [Configuration] Screen