Triggering Examples
Cross-Arming Trigger Examples
To trigger timing analysis of a count-down on a
set of data lines
Your target system may include various state machines that are started
by system events such as interrupt processing or I/O activity. The state
analyzer is ideal for recognizing the system events; the timing analyzer
is ideal for examining the step-by-step operation of the state machines.
1 Set up a timing analyzer and a state analyzer.
2 Go to the state analyzer's Trigger menu.
3 Set the timing analyzer to be run from the state analyzer.
4 Set the state analyzer to trigger on the label and term that
identify the start of the count-down routine.
5 Go to the timing analyzer's Trigger Menu.
6 Set the timing analyzer to trigger on any state and store any
state.