DisplayPort Test Implementation Group 14 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Figure 5: J-BERT N4903B HBR2 Initial Amplitude
Figure 6: J-BERT N4903B HBR2 TP3_EQ ISI Verification
Note: In this example, the measured ISI value is 220.1mUI, small variations between the
measured value and the target value of jitter will be compensated for in 9.3.2. Capture
and save the screenshot of the oscilloscope measurement to demonstrate that the proper
value has been achieved.
Important: In order to setup the oscilloscope for best measurements, follow the procedure
in B appendix. It is important to get the vertical signals properly scaled without clipping
before applying Ch1-Ch3 function and equalization.