Agilent Technologies DisplayPort Manual de usuario Pagina 35

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 34
Agilent Technologies, Inc.
DisplayPort Test Implementation Group 35 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
k. Press Close to leave the Jitter menu.
l. The DSA will take a few moments to calculate and display the Jitter measurement results.
Figure 32: HBR2 Differential TP3-EQ Signal with ISI Measurement
m. Verify ISI is within 5% of the specification limit. If it is not 5%, adjustments on the ISI
channel are needed to be within a +/- 5% target. Adjust channel and try again. To acquire a
Vista de pagina 34