DisplayPort Test Implementation Group 31 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Figure 29: HBR2 Equalizer Menu Settings
b. For HBR:
i. Data Rate = 2.7Gb/s
ii. Equalizer=CTLE
iii. # of Poles = 2
iv. DC Gain = 1.0
v. Zero Frequency = 725MHz
vi. Pole 1 Frequency = 1.35GHz
vii. Pole 2 Frequency = 2.5GHz
c. Press Close on Equalizer Setup Menu.
d. Turn Off C1-C3 Waveform so only Equalized Waveform is displayed.
3. The results in the TP3_EQ waveform for HBR2 and HBR and non-equalized waveform for RBR.