Electrode Cartridge Tests
Fan Test Checks if the fan is running at the appropriate speed.
Temperature Test Checks if the temperature ramp-up speed of the heater plate is
within specifications.
Diagnostics Test Purpose
Current Leakage Test Measures electrode cartridge leak current(s) between pins.
Leak current test chip required. For correct preparation of the
leak current test chip, please refer to the technical note
included in the test chip bundle.
Short Circuit Test Checks for instrument leak currents using an empty chip.
Note: the limits of this test specify an ambient temperature of
25 °C and relative humidity less than or equal to 60 %. Higher
temperatures or relative humidity could result in a leak current.
Optics Test Checks for proper alignment of internal optics and proper
function of the laser and LED.
High-Voltage Stability
Te st
Tests high voltage accuracy and stability of all 16 high voltage
power supplies. Unused chip (DNA, RNA, or protein) required.
High-Voltage Accuracy
Te st
Check of the high voltage controller.
Diagnostics Test Purpose