Contents 49 Index
How to Prepare the Bioanalyzer for Electrophoretic Assays
Remove the pressure cartridge:
1 Turn off line power to the Agilent 2100 bioanalyzer. The line switch is located at the rear.
2 Open the lid.
3 Pull down the metal locking lever as shown in the figure below.
The cartridge is pushed out.
4 Gently pull the cartridge out of the lid.
NOTE
Store the pressure cartridge in the provided box.
Metal lever
in open
position