Contents 314 Index
Validation Procedure
To perform qualification tests:
1 Switch to the Validation context.
2 From the File menu select New.
3 A New Validation item appears in the Tree View Panel.
4 Under Cartridge Details, click on Select... and specify details on the cartridge that is
currently installed in the bioanalyzer.
5 Under Configure 2100 Bioanalyzer HW Test Chips, enter the test chips you will use for
this validation: