2 MEASurement Commands
104 N1911A/1912A P-Series Power Meters Programming Guide
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence:
RELative [<expected_value>[,<resolution>[,<source list>]]]
This command sets the measurement function, range and resolution of the
specified window. It sets the measurement function to difference with
relative mode on. The relative value used is set by the
CALCulate:RELative:MAGNitude:AUTO command.
Syntax
Parameters
Refer to “Optional Parameters” on page 90 for additional details on the
parameters in this command.
:POW
:AC
:SCAL
:DIFF
:REL
Space
expected_value
DEF
resolution
,
DEF
source list
,
CONF
1
2
3
4
Item Description/Default Range of Values
expected_value
(for the expected
power level)
The power meter ignores the numeric
value entered in this parameter. Any
value entered is treated like DEF.
sensor dependent
DEF
1
resolution A numeric value for the resolution. If
unspecified the current resolution setting
is used.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1