7MEMory Subsystem
276 N1911A/1912A P-Series Power Meters Programming Guide
MEMory:STATe:DEFine <character_data>,<numeric_value>
This command is used to associate a name with a save/recall register
number.
Syntax
Parameters
Example
Query
MEMory:STATe:DEFine? <string>
The query returns the register number for the given register name.
Item Description/Default Range of Values
character_data Details the register name. A maximum of 12
characters can be used.
A to Z (uppercase)
a to z (lowercase)
0-9
_ (underscore)
numeric_value A numeric value (<NRf>) for the register
number.
0 to 9
MEM:STAT:DEF "SETUP1",4 This command names register 4 SETUP1.
MEM
:STAT
numeric_value
:DEF
,
Space
?
Space
character_data
character_data