10 SENSe Subsystem
374 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:AVERage2:COUNt <numeric_value>
This command is used to enter the video filter length for the P- Series and
E9320 sensor. Video filtering is applied to the traces. Successive traces are
combined to reduce noise without affecting the dynamic characteristic of
the signal.
Syntax
Parameters
1
This is only implemented in powers of 2 (2
n
).
Example
Item Description/Default Range of Values
numeric_value A numeric value defining the filter length.
DEF: the default value is 4.
1 to 256
1
DEF
:COUN
Space
?
numeric_value
DEF
SENS
1
AVER2
SENS2
:
AVER2:COUN 16 This command enters a video filter length
of 16 for Channel A.