9 PSTatistic Subsystem
332 N1911A/1912A P-Series Power Meters Programming Guide
PSTatistic[1]|2:CCDF:COUNt <numeric_value>
This command is used to set CCDF cummulative counts for Channel A or
Channel B. The increment step for the cummulative counts is 100 M.
Syntax
Parameters
Example
NOTE
This command is only applicable when P-Series sensor is present and free-run acquisition
mode is selected. If P-Series sensor is used by the adjacent channel in a dual channel
setup, the same setting will be applied for free-run acquisition mode.
:CCDF :COUN
numeric_value
Space
?
PST
2
1
Item Description/Default Range of Values
numeric_value The CCDF cummulative counts in numeric
value.
Minimum value: 100 M
Maximum value: 10 G
100 M to 10 G
PST1:CCDF:COUN 1.2G This command sets the CCDF
cummulative counts for Channel A to 1.2
G.