MEMory Subsystem 7
N1911A/1912A P-Series Power Meters Programming Guide 287
MEMory:TABLe:GAIN[:MAGNitude]:POINts?
This query is used to return the number of calibration factor/offset points
for the currently selected table. If the currently selected table is a sensor
calibration table (8480 Series sensors and N8480 Series sensors with
Option CFT only), the reference calibration factor is included
If no values have been set, 0 is returned. If no table is selected, NAN is
returned.
Syntax
Example
MEM:TABL:GAIN:POIN? This command queries the number of
calibration factor/offset points in the
current table.
MEM
:TABL
:GAIN
:MAGN
:POIN
?