11 STATus Subsystem
476 N1911A/1912A P-Series Power Meters Programming Guide
STATus:OPERation
The operation status register set contains conditions which are a part of
the operation of the power meter as a whole.
The following bits in these registers are used by the power meter:
Syntax
Bit Number Decimal
Weight
Definition
0 1 CALibrating Summary
1 - 3 - Not used
4 16 MEASuring Summary
5 32 Waiting for TRIGger Summary
6 - 9 - Not used
10 1024 SENSe Summary
11 2048 Lower Limit Fail Summary
12 4096 Upper Limit Fail Summary
13 to 15 - Not used (bit 15 always 0)
STAT
:OPER