Agilent Technologies 1912A Manual de servicio Pagina 735

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 803
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 734
SERVice Subsystem 16
N1911A/1912A P-Series Power Meters Programming Guide 707
SERVice:BIST:TBASe:STATe <boolean>
This command sends a 10 MHz time base signal to the rear panel trig out
for testing purposes.
If the command is set to:
ON, the 10 MHz time base signal is sent to the rear panel trigger out
connector.
OFF, the 10 MHz time base signal is disabled.
Syntax
Example
NOTE
This command overrides the OUTPut:TRIGger[:STATe] command.
For example, if OUTPut:TRIGger[:STATe] is ON and the command
SERV:BIST:TBAS ON is sent, this command overrides the Trigger state and sets it to
OFF. However, the 10 MHz remains out the Trig out port.
If the SERV:BIST:TBAS ON has been sent, the 10 MHz is on and the
OUTPut:TRIGger[:STATe] is then toggled to ON, the channel trigger is now routed to
the Trig out overriding the service command turning the 10 MHz to off.
SERV:BIST:TBAS:STAT OFF This command disables the signal.
:BIST
SERV
0|OFF
1|ON
Space
?
:TBAS
:STAT
Vista de pagina 734
1 2 ... 730 731 732 733 734 735 736 737 738 739 740 ... 802 803

Comentarios a estos manuales

Sin comentarios