3CALCulate Subsystem
190 N1911A/1912A P-Series Power Meters Programming Guide
Parameters
1
Quotes are mandatory. Either single or double quotes may be used.
2
N1912A only.
3
The mathematical operation will be performed in linear scale.
Example
Reset Condition
On reset, the Agilent N1911A upper and lower window measurements are
set to Channel A ("(SENS1)"). On the N1912A the upper window
measurements are set to Channel A ("(SENS1)") and the lower window
measurements to Channel B ("(SENS2)").
Item Description/Default Range of Values
string A single string value detailing the measurement
type:
For the Agilent N1911A the default is
SENS1.
For the Agilent N1912A the default is
SENS1 if the upper window is selected, or
SENS2 if the lower window is selected.
“(SENS1)”
1
“(SENS2)”
1,2
“(SENS1–SENS1)”
1,3
“(SENS2–SENS2)”
1,2,3
“(SENS1/SENS1)”
1
“(SENS2/SENS2)”
1,2
“(SENS1–SENS2)”
1,2,3
“(SENS2–SENS1)”
1,2,3
“(SENS1/SENS2)”
1,2
“(SENS2/SENS1)”
1,2
CALC2:MATH “(SENS2/SENS1)” This command sets the lower
window/upper measurement to make a
Channel B/A ratio measurement.