10 SENSe Subsystem
408 N1911A/1912A P-Series Power Meters Programming Guide
Parameters
Example
Reset Condition
On reset, GAIN2 is set to 0.00 dB.
Query
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude]?
[MIN|MAX]
The query returns the current setting of the channel offset or the values
associated with MIN and MAX.
Item Description/Default Range of Values
numeric_value A numeric value:
DEF: the default is 0.00 dB
MIN: –100 dB
MAX: +100 dB
–100 to +100 dB
DEF
MIN
MAX
CORR:GAIN2 50 This command sets a channel offset of
50 dB for Channel A.