102 Chapter 1
GPIB Command Dictionary
CANalyzer subsystem
:CHANnel:POWer:FILTer
:CHANnel:POWer:FILTer?
These commands set/query the filter for the channel power
measurement.
‘1.23M’ measures the absolute channel power level of the RF signal
in a 1.23 MHz bandwidth centered around the RF channel or tune
frequency setting. This measurement must be calibrated using the
CAN:CHAN:CPOW:CAL command each time the frequency is
changed.
‘30kHz’ measures the adjacent channel power (that is the power in
the sidebands around the chosen CDMA channel).
Syntax
CAN:CHAN:POW:FILT ‘1.23M’
‘30kHz’
CAN:CHAN:POW:FILT?
Screen/field equivalent
CAN:CHAN:POW:FILT controls the Ch Pwr Filt field of the CDMA
ANALYZER screen while measuring channel power.
:DARM
This command works with the :ARM command to control the
measurements.
Syntax
CAN:DARM
Screen/field equivalent
CAN:DARM controls the Disarm subfield of the Analyzer field on the
CDMA ANALYZER screen.