Agilent Technologies 8935 Series Manual de usuario Pagina 96

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 352
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 95
96 Chapter 1
GPIB Command Dictionary
AFG2|ENCoder subsystem
:NAMPs|NTACs:FVC:RATE <real number>
:NAMPs|NTACs:FVC:RATE?
These commands set/query the data rate of the encoded message.
This command utilizes the “Real Number Setting Syntax” on page 299
but does not use the :STATe command.
Syntax
AFG2:NAMP:FVC:RATE <real number>
AFG2:NAMP:FVC:RATE?
Screen/field equivalent
AFG2:NAMP:FVC:RATE controls the Data Rate field on the
SIGNALING ENCODER screen when the Mode field is set to
NAMP-NTAC, and the Channel field is set to Voice.
:NAMPs|NTACs:FVC:AM <real number>
:NAMPs|NTACs:FVC:AM?
This command sets/queries the data level when the AFG2:DEST ‘AM’
command is used. The data level units are %.
This command utilizes the “Real Number Setting Syntax” on page 299.
Syntax
AFG2:NAMP:FVC:AM
AFG2:NAMP:FVC:AM?
Screen/field equivalent
AFG2:NAMP:FVC:AM controls the Data Level field when AFGen2 To
field is set to AM on the SIGNALING ENCODER screen when the Mode
field is set to NAMP-NTAC, and the Channel field is set to Voice.
Vista de pagina 95