Chapter 1 113
GPIB Command Dictionary
CANalyzer subsystem
:TRIGger:STATe <string>
:TRIGger:STATe?
These commands set/query the state of the trigger.
CAN:TRIG:STAT can be overridden with the TRIG:MODE:RETR REP
and TRIG:MODE:RETR SING trigger commands. When the trigger
commands are used, the state of the field on the display is not updated.
(For instance, the screen may show Single selected when
TRIG:MODE:RETR REP is used.)
Syntax
CAN:TRIG:STAT ‘Single’
‘Cont’
CAN:TRIG:STAT?
Screen/field equivalent
CAN:TRIG:STAT controls Single/Cont subfield of the Analyzer field
on the CDMA ANALYZER screen.