Chapter 1 201
GPIB Command Dictionary
DECoder subsystem
:DTMF:GATE <real number>
:DTMF:GATE?
This command specifies how long the decoder analyzes a signal after it
has been triggered. Up to 6.55 seconds of gate time can be specified.
If the gate time is too long, the decoders buffer becomes full. A message
is displayed instructing you to decrease the gate time.
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STATe command.
Syntax
DEC:DTMF:GATE <real number>
DEC:DTMF:GATE?
Screen/field equivalent
DEC:DTMF:GATE controls the Gate Time field on the SIGNALING
DECODER screen when the Mode field is set to DTMF.
:FGEN:GATE <real number>
This command specifies how long the decoder analyzes a signal after it
has been triggered. Up to 6.55 seconds of gate time can be specified.
If the gate time is too long, the decoders buffer becomes full. A message
is displayed instructing you to decrease the gate time.
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STATe command.
Syntax
DEC:FGEN:GATE <real number>
Screen/field equivalent
DEC:FGEN:GATE controls the Gate Time field on the SIGNALING
DECODER screen when the Mode field is set to Func Gen.