Chapter 1 227
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:CPOWer <meas cmnd>
:CDANanlyzer:CPOWer?
IS-95 Only mode. These command set/query the channel power
measurement.
This measurement utilizes the Number Measurement Syntax, but does
not use the :METer command.
Syntax
MEAS:CDAN:CPOW <meas cmd>
!See “Number Measurement Syntax” on page 283.
MEAS:CDAN:CPOW?
Screen/field equivalent
MEAS:CDAN:CPOW? reads the ChPwr measurement field on the CD
Setup menu of the CODE DOM screen when the CDMA standard is set
to “IS-95 Only”.
:CDANalyzer:ERRor:FREQuency <meas cmnd>
:CDANalyzer:ERRor:FREQuency?
IS-95 Only mode. These commands set/query the frequency error
measurement on the code domain analyzer.
This measurement utilizes the Number Measurement Syntax, but does
not use the :METer command.
Syntax
MEAS:CDAN:ERR:FREQ <meas cmd>
!See “Number Measurement Syntax” on page 283.
MEAS:CDAN:ERR:FREQ?
Example
MEAS:CDAN:ERR:FREQ?
Screen/field equivalent
MEAS:CDAN:ERR:FREQ? reads the Freq Err measurement field on
the CODE DOM screen when the CDMA standard is set to “IS-95
Only”.