132 Chapter 1
GPIB Command Dictionary
CDANalyzer subsystem
:IS2000:GAIN:MODE <string>
:IS2000:GAIN:MODE?
For IS-2000 only. These commands set/query the mode of the Gain field,
allowing you to choose between an autoranging gain (‘Auto’) or a fixed
gain (‘Hold’).
This command may conflict with the CDMA analyzers settings
CAN:PATH:GAIN and CAN:PATH:MODE. For example, if autoranging
is turned off (‘Hold’) in the CDMA analyzer and turned on in the Code
Domain Analyzer (‘Auto’), the ‘Hold’ setting will always override the
‘Auto’ setting. When autoranging is turned off, setting either analyzers
path gain will automatically change the others to the same setting.
For average power measurements, use the CAN:POW:GAIN:MODE
command.
Syntax
CDAN:IS2000:GAIN:MODE ‘Auto’
‘Hold’
CDAN:IS2000:GAIN:MODE?
Screen/field equivalent
CDAN:IS2000:GAIN:MODE sets the Auto/Hold subfield of the Gain
image of the CODE DOMAIN screen when the CDMA standard is set to
“IS-2000”.
:IS2000:GAIN:PATH <real number>
:IS2000:GAIN:PATH?
For IS-2000 only. These commands query/set returns the gain
information from the Code Domain analyzer. The command allows you
to fine tune the gain setting.
Syntax
CDAN:IS2000:GAIN:PATH <real number> !values from 0 to 36 dB
in 0.1 dB steps
CDAN:IS2000:GAIN:PATH?
Screen/field equivalent
CDAN:IS2000:GAIN:PATH has no corresponding field. This applies
when the CDMA standard is set to “IS-2000”.