144 Chapter 1
GPIB Command Dictionary
CDANalyzer subsystem
:IS2000:TRIGger:DELay <real number>
:IS2000:TRIGger:DELay?
IS-2000 only. This command/query sets/returns the real number that
matches the Trig Delay field in the code domain measurements. The
contents are measured in microseconds.
Syntax
CDAN:IS2000:TRIG:DEL <real number> !values from 20µs to 10 s
CDAN:IS2000:TRIG:DEL?
Screen/field equivalent
CDAN:IS2000:TRIG:DEL controls the Trig Delay subfield of the
Trigger image when the Trig Event is set to Delay on the CODE
DOMAIN screen when the CDMA standard is set to “IS-2000”.
:IS2000:TRIGger[:EVENt] <string>
:IS2000:TRIGger[:EVENt]?
IS-2000 only. This command returns the string that matches the Trig
Event field in the code domain measurements.
Syntax
CDAN:IS2000:TRIG[:EVEN] ‘27 ms’
‘20 ms’
‘80 ms’
‘2 s’
‘Delay’
‘Immed’
CDAN:IS2000:TRIG[:EVEN]?
Screen/field equivalent
CDAN:IS2000:TRIG[:EVEN] controls the Trig Event field of the
Trigger image of the CODE DOMAIN screen when the CDMA
standard is set to “IS-2000”.