Agilent Technologies 8510C Manual de servicio Pagina 102

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 222
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 101
3-32 8510XF Network Analyzer Systems
Operation
RF Power
Choices
If {DETECT UNL / ALWAYS} is selected, the 8510C polls for errors during every
sweep.
If
{DETECT UNL / SMART} is selected, the 8510C polls for errors during the first
sweep following a frequency change, and thereafter only if an error was
detected during the first sweep. This is the default mode.
If
{DETECT UNL / ONCE} is selected, the 8510C polls for errors only during the
first sweep following a frequency change.
If
{DETECT UNL / NEVER} is selected, the 8510C does not poll for errors during
any sweep.
Blanking On/Off The softkeys {BLANKING /ON} and {BLANKING /OFF} activate and deactivate port
power blanking during frequency or port drive transitions. (See “Blanking
circuit” on page 5-17).
Reset Detector Gain
Calibration
The {RESET DET GAIN CAL} softkey calls up a submenu (see “Reset Detector
Gain Calibration” on page 3-34).
Vista de pagina 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 221 222

Comentarios a estos manuales

Sin comentarios