8510XF Network Analyzer Systems 3-35
Operation
RF Power
“RUN CAL” Functions If {RUN CAL + USER PRST} is selected, the detector gain calibration routine is
run, and is followed by a user preset.
If
{RUNCAL+FACT PRST} is selected, the detector gain calibration routine is run,
and is followed by a factory preset.
If
{RUN CAL NO PRESET} is selected, the detector gain calibration routine is run,
but is not followed by a preset (system settings that are altered by the
calibration routine will not be restored to their original conditions
afterward).
Use
{CANCEL} to exit the menu without running the calibration routine.