8510XF Network Analyzer Systems 3-25
Operation
Standard Types
One-Path Isolation
Standard
This standard is used to measure forward isolation for one-path 2-port
calibrations. When you perform a calibration of that type using a 1.0 mm
calibration kit, you will see the following menu:
Figure 3-13 One-Path Isolation Calibration Menu & Equivalent GPIB Commands
Note on the Standard {FWD ISOL’N
LOAD}
Refer to “Isolation” under “Banded Standards” in
Table 3-1 on page 3-12.
NOTE To skip the isolation part of the calibration, select {OMIT ISOLATION}.